Acquisition Phase

Phase Included

Scan Delay

Respiration Phase

Anatomical Coverage

Non Contrast

Yes

N/A

Inspiration

Carina through the apex of the heart

Arterial Phase

N/A

N/A

N/A

N/A

Venous Phase

N/A

N/A

N/A

N/A

Delayed Phase

N/A

M/A

N/A

N/A

Scan Comments: Flash spiral mode is used for calcium scoring on all patients with heart rates up to 70 bpm. For patients with heart rates greater than 70 bpm, prospective sequential imaging is used instead. All scans are performed using the ALARA principle.

 

Technical Parameters

Parameters

kVp

Semi (ref. kV 120)

Effective mAs

CareDose

Quality Reference mAs

80

Time (Rotation)

0.28 sec.

Average Acquisition Time

1-2 sec.

Collimation

128 x 0.6 mm

Pitch Value

3.4

Scan Direction

Craniocaudal

Comments: CareDose is used to reduce patient dose.

 

Reconstruction Parameters

Soft Tissue

Soft Tissue Open Field

Thin Data Open Field

Lung

Slice Thickness

 3 mm

3 mm

 0.75 mm

3 mm

Reconstruction Spacing

 3 mm

3 mm

 0.5 mm

3 mm

Reconstruction Algorithm

Qr36

Qr36

Qr36

Bl57

Window Width and Level

350/50

410/10

410/10

1600/-500

Reconstruction Comments: The reconstruction field of view is limited to 33.2cm using the flash spiral scan mode.

 

Contrast Parameters

Parameters

Contrast Type

N/A

Contrast Volume

N/A

Saline Flush

N/A

Injection Rate

N/A

Oral Contrast

N/A

Contrast Volume

N/A

Comments:IV contrast is not necessary for this exam.