Acquisition Phase

Phase Included

Scan Delay

Respiration Phase

Anatomical Coverage

Non Contrast

Yes

N/A

N/A

Base of skull through vertex

Arterial Phase

 N/A

N/A

N/A

N/A

Venous Phase

 N/A

N/A

N/A

N/A

Delayed Phase

 N/A

N/A

N/A

N/A

Scan Comments:Flash spiral is very useful when imaging patients who are unable to hold still for a routine spiral study to evaluate boney skull pathology. This protocol should not be used to evaluate soft tissue structures. It should be noted that image quality will be slightly degraded do to the high pitch value; however, patient motion will be significantly limited. All scans are performed using the ALARA principle.

 

Technical Parameters

Parameters

kVp

Care kV (ref. kV 120)

Effective mAs

CareDose

Care Dose Reference mAs

210

Time (Rotation)

0.28 sec

Average Acquisition Time

2-3 sec

Collimation

128 x 0.6mm

Pitch Value

1.55

Scan Direction

Caudocranial

Comments: CareDose is used to reduce patient dose.

 

 

Reconstruction Parameters

Soft Tissue

Bone

Soft Tissue

Thin Data

Slice Thickness

5 mm

5 mm

5 mm

0.75 mm

Reconstruction Spacing

5 mm

5 mm

5 mm

0.5 mm

Reconstruction Algorithm

H30f

H60f

J30f (3)

H30f

Window Width and Level

100/30

2500/500

100/30

100/30

Reconstruction Comments: 3D MPR’s can be performed to correct positioning.

 

Reconstruction Parameters

Thin Data

Thin Data

Coronal

Sagittal

Slice Thickness

0.75 mm

0.75 mm

3 mm

3 mm

Reconstruction Spacing

0.5 mm

0.5 mm

3 mm

3 mm

Reconstruction Algorithm

J30f (3)

H60f

H60f

H60f

Window Width and Level

100/30

2500/500

2500/500

2500/500

Reconstruction Comments: 3D MPR’s can be performed to correct positioning.

 

Contrast Parameters

Parameters

Contrast Type

 N/A

Contrast Volume

 N/A

Saline Flush

 N/A

Injection Rate

 N/A

Oral Contrast

 N/A

Contrast Volume

 N/A

Comments: IV contrast is not typically necessary for this exam.