Acquisition Phase

Phase Included

Scan Delay

Respiration Phase

Anatomical Coverage

Non Contrast

Yes

N/A

N/A

Maxillary sinus through frontal sinus

Arterial Phase

N/A

N/A

N/A

N/A

Venous Phase

N/A

N/A

N/A

N/A

Delayed Phase

N/A

N/A

N/A

N/A

Scan Comments: Iso-centering the patient within the gantry is key to image quality. All scans are performed using the ALARA principle.

 

Technical Parameters

Parameters

kVp

Care kV (ref. kV 120)

Effective mAs

CareDose

Quality Reference mAs

150

Time (Rotation)

1.0 sec

Average Acquisition Time

5-7 sec

Collimation

128 x 0.6mm

Pitch Value

0.8

Scan Direction

Caudocranial

Comments: CareDose is used to reduce patient dose.

 

Reconstruction Parameters

Bone

Soft Tissue

Thin Data

Thin Data

Slice Thickness

3 mm

3 mm

0.75 mm

0.75 mm

Reconstruction Spacing

3 mm

3 mm

 0.5 mm

0.5 mm

Reconstruction Algorithm

J70h (3)

J30s (3)

J70h (3)

J30s (3)

Window Width and Level

2500/500

350/20

2500/500

350/20

Reconstruction Comments: 

 

Reconstruction Parameters

Coronal

Sagittal

Coronal

Sagittal

Slice Thickness

2 mm

2 mm

2 mm

2 mm

Reconstruction Spacing

2 mm

2 mm

2 mm

2 mm

Reconstruction Algorithm

J70h (3)

J70h (3)

J30s (3)

J30s (3)

Window Width and Level

2500/500

2500/500

350/20

350/20

Reconstruction Comments: 

 

Contrast Parameters

Parameters

Contrast Type

N/A

Contrast Volume

N/A

Saline Flush

N/A

Injection Rate

N/A

Oral Contrast

N/A

Contrast Volume

N/A

Comments: IV contrast is not typically necessary for this exam.