• Prospective-triggered sequential dual-source end-systolic coronary CT angiography for patients with atrial fibrillation: A feasibility study.

    J Cardiovasc Comput Tomogr. 2013 Mar-Apr;7(2):102-9. doi: 10.1016/j.jcct.2013.02.002. Epub 2013 Feb 27.

    Srichai MB, Barreto M, Lim RP, Donnino R, Babb JS, Jacobs JE.

    Source

    Department of Medicine, Cardiology Division, Medstar Georgetown University Hospital, 3800 Reservoir Road NW, 5PHC, Washington, DC 20007, USA. Electronic address: [email protected].

    Abstract

    BACKGROUND:

    Obtaining diagnostic coronary CT angiography with low radiation exposure in patients with irregular heart rhythms such as atrial fibrillation (AF) remains challenging.

    OBJECTIVE:

    We evaluated image quality and inter-reader variability with the use of prospective electrocardiographic (ECG)-triggered sequential dual-source acquisition at end systole for coronary artery disease (CAD) evaluation in patients with AF.

    METHODS:

    Thirty consecutive patients with AF who underwent prospective ECG-triggered sequential dual-source acquisition were evaluated. Images were reconstructed every 50 milliseconds from 250 to 400 milliseconds after the R wave. Two independent, blinded readers evaluated the coronaries for image quality on a 5-point scale (worst to best) and stenosis on 5-point semiquantitative (none to severe) and binary scales (>50% or <50%). Diagnostic image quality was graded for each reconstruction.

    RESULTS:

    Eleven patients (37%) had significant (≥50% stenosis) CAD. Average heart rate was 82 ± 20 beats/min and variability range was 71 ± 22 beats/min. Mean effective radiation dose was 6.5 ± 2.4 mSv. Diagnostic image quality was noted in 97.9% of 304 coronary segments with median image quality of 3.0. The 300-millisecond reconstruction phase provided the highest image quality; 70% of patients showed diagnostic image quality. Combination of all phases (250-400 milliseconds) performed significantly better than single or other phase combinations (P < 0.0005 for all comparisons). Inter-reader variability for stenosis detection was excellent, with 98.4% concordance by using a binary scale (50% stenosis cutoff).

    CONCLUSIONS:

    Prospective ECG-triggered sequential dual-source CT acquisition with the use of end-systolic acquisition provides diagnostic image quality with potentially low radiation doses for evaluation of CAD in patients with AF. Use of multiple end-systolic phases over a 150-millisecond window improves diagnostic image quality.