• Quantitative Assessment of Lung Cancer Perfusion Using MDCT: Does Measurement Reproducibility Improve with Greater Tumor Volume Coverage?

    AJR 2006; 187:1079-1084.

    Ng QS, Goh V, Klotz E, Fichte H, Saunders MI, Hoskin PJ, Padhani AR.

    OBJECTIVE: To date, quantitative assessment of tumor vascularity using perfusion CT has been limited to a single tumor level, with the potential for measurement error in heterogeneous tumors. We aimed to determine if greater z-axis tumor coverage improves the reproducibility of perfusion CT measurements in lung cancer.

    SUBJECTS AND METHODS: Paired perfusion studies were performed on 10 patients who had histologically confirmed advanced non-small cell lung cancer. Using 16-MDCT, multiple sequential volumetric acquisitions encompassing the entire tumor were acquired after infusion of IV contrast material. Using Patlak analysis, median values of tumor permeability (mL/100 mL/min) and blood volume (mL/100 mL) were measured for 10-mm z-axis coverage, and for 40-mm z-axis coverage in each of the paired perfusion studies. Measurement reproducibility was evaluated using Bland-Altman statistics.

    RESULTS: Mean difference (95% limits of agreement) for tumor permeability was 1.4 (-4.0 to 6.8) for 10-mm coverage and 0.8 (-3.6 to 5.2) for 40-mm coverage. Mean difference (95% limits of agreement) for blood volume was 1.9 (-5.1 to 8.9) for 10-mm coverage and 1.4 (-3.7 to 6.6) for 40-mm coverage. The coefficient of variation for permeability was 18.7% for 10-mm coverage, improving to 11.9% for 40-mm coverage. The coefficient of variation for blood volume was 41.7% for 10-mm coverage, improving to 32.6% for 40-mm coverage.

    CONCLUSION: Our results show that an improvement in tumor perfusion measurement reproducibility may be achieved with greater z-axis coverage.