Objective: The purpose of this study was to compare prospective ECG-triggered and retrospective ECG-gated coronary 64-MDCT angiography as to radiation dose, image quality, accuracy of stenosis measurement, and CT densitometry.
Materials and Methods: Coronary artery models (n = 3) with different plaque densities (= 50, = 110, and = 1,000 H) on a cardiac phantom were scanned in variable heart rate sequences (n = 14) with both prospective ECG-triggered and retrospective ECG-gated scanning. Radiation dose, image quality graded by motion and stairstep artifacts (grade 1, excellent, to grade 4, poor, with grades 1 and 2 defined as satisfactory), accuracy of stenosis measurement (area; 18%, 50%, and 82%), and CT densitometry of plaques (soft, = 50; and intermediate, " 110 H) were compared between the two protocols using the Mann-Whitney U test and repeated measures.
Results: The radiation dose of prospective ECG-triggered CT angiography (CTA) (3.0 mSv) was lower than that of retrospective ECG-gated CTA (11.7-13.0 mSv) when the same tube current (mA) and voltage (kVp) were used in both methods. Prospective ECG-triggered CTA images were assigned a satisfactory quality rating in stable heart rate up to 75 beats per minute (bpm) when using the minimal X-ray exposure time. In this range, there were no significant differences in stenosis measurement (p = 0.17) and CT densitometry (p = 0.93) between the two protocols.
Conclusion: Prospective ECG-triggered coronary 64-MDCT has the potential to reduce radiation exposure while maintaining the diagnostic performance of retrospective ECG-gated coronary 64-MDCT.