EP No: 14

Scanogram - Lateral Scanogram Sternum to mid head
Scan modeScanogram(Lat)
Start time(s)0
Wait time(s)0
kV120
mA200
Range(mm)500
DirectionOUT
Display filterSharp (FL03)
Scano angle °90
CEOFF

 

Scanogram - AP Scanogram
Scan modeScanogram(AP)
Start time(s)0
Wait time(s)6.5
kV120
mA50
Range(mm)500
DirectionIN
Display filterSharp (FL03)
Scano angle °0
CEOFF

 

S&V - Position below base of skull Manual trigger
Scan modeS&V
# of scan(s)1
Start time(s)***
Wait time(s)0
Collimation0.5 x 4
Pitch***
kV120
mA50
Rotation time(s)0.5
Total scan time(s)0.5
Range(mm)2
DirectionOUT
CFOVM
CEOFF
CTDI(mGy)6.8
DLP(mGy.cm)1.4

 

ReconstructionAxial1
Slice thickness2.0 mm
Slice Interval***
DFOV260
SUREIQNeck - Low Dose
Recon ProcessOFF
FC7
Boost 3DOFF
OSR***
FilterOFF

 

SureStart - Manual trigger as soon as contrast is seen
Scan modeSureStart
# of scan(s)***
Start time(s)***
Wait time(s)10
Collimation0.5 x 4
Pitch***
kV120
mA50
Rotation time(s)0.5
Total scan time(s)40
Range(mm)0
Direction***
CFOVM
CEON
CTDI(mGy)523.9
DLP(mGy.cm)104.8

 

ReconstructionAxial1
Slice thickness2.0 mm
Slice Interval***
DFOV260
SUREIQNeck - Low Dose
Recon Process***
FC7
Boost 3DOFF
OSR***
FilterOFF

 

Helical - Mid head to below origins of arch vessels
Scan modeHelical
# of scan(s)1
Start time(s)0
Wait time(s)0
Collimation0.5 x 80
Pitch69
kV120
mAExp.3D Quality
Rotation time(s)0.5
Total scan time(s)4.5
Range(mm)252
DirectionIN
CFOVM
CEON
CTDI(mGy)***
DLP(mGy.cm)***
ReconstructionAxial1Volume1
Slice thickness3.0 mm0.5 mm
Slice Interval3.0 mm0.30 mm
DFOV260260
SUREIQCTA - NeckCTA - Neck
Recon ProcessAIDR 3D StandardAIDR 3D Standard
FC4343
Boost 3DOFFOFF
OSRONON
FilterOFFOFF

 

* Exp.3D Adult Head Quality Preset - Reference Image Quality
Target SD10
Target Slice thickness3.0 mm
Target SUREIQNeck Standard
Recon ProcessAIDR 3D Standard
FC8
XY modulationON
Max(mA)700
Min(mA)100