Neck HCT 3mm

EP No: 48

Scanogram - Lateral Scanogram Below base of skull to vertex
Scan modeScanogram(Lat)
Start time(s)0
Wait time(s)0
kV120
mA150
Range(mm)500
DirectionOUT
Display filterSharp (FL03)
Scano angle °90
CEOFF

 

Scanogram - AP Scanogram
Scan modeScanogram(AP)
Start time(s)0
Wait time(s)6.1
kV120
mA50
Range(mm)500
DirectionIN
Display filterSharp (FL03)
Scano angle °0
CEOFF

 

Helical - Arterial neck Above pit fossa to aortic arch
Scan modeHelical
# of scan(s)1
Start time(s)40
Wait time(s)40
Collimation0.5 x 80
PitchDetail
kV120
mAExp.3D Quality
Rotation time(s)0.275
Total scan time(s)3.9
Range(mm)300
DirectionIN
CFOVM
CEON
CTDI(mGy)***
DLP(mGy.cm)***
ReconstructionAxialVolume 1MutiView1 CoronalMutiView1 Sagittal
Slice thickness3.0 mm0.5 mm3.0 mm3.0 mm
Slice Interval3.00 mm0.30 mm3.00 mm3.00 mm
DFOV280280******
SUREIQNeck - StandardNeck - StandardNeck - StandardNeck - Standard
Recon ProcessAIDR 3D StandardAIDR 3D Standard******
FC88******
Boost 3DOFFOFF******
OSRONON******
FilterOFFOFF******

 

* Exp.3D Adult Neck Quality Preset - Reference Image Quality
Target SD10
Target Slice thickness3.0 mm
Target SUREIQNeck Standard
Recon ProcessAIDR 3D Standard
FC8
XY modulationON
Max(mA)700
Min(mA)100