EP No: 104

Scanogram - AP Scanogram Above shoulders to below liver
Scan modeScanogram(AP)
Start time(s)0
Wait time(s)0
kV120
mA50
Range(mm)500
DirectionOUT
Display filterSharp (FL03)
Scano angle °0
CEOFF

 

Scanogram - Lateral Scanogram
Scan modeScanogram(Lat)
Start time(s)0
Wait time(s)6.1
kV120
mA100
Range(mm)500
DirectionIN
Display filterSharp (FL03)
Scano angle °90
CEOFF

 

S&V - Mid volume
Scan modeS&V
# of scan(s)1
Start time(s)***
Wait time(s)0
Collimation1.0 x 4
Pitch***
kV100
mA80
Rotation time(s)0.275
Total scan time(s)0.275
Range(mm)4
DirectionOUT
CFOVM
CEOFF
CTDI(mGy)2.4
DLP(mGy.cm)1

 

ReconstructionAxial1
Slice thickness4.0 mm
Slice Interval***
DFOV220
SUREIQCardiac - Low Dose
Recon ProcessOFF
FC2
Boost 3DOFF
OSR***
FilterOFF
SureStart - Position in descending aorta Trigger at 180 HU
Scan modeSureStart
# of scan(s)***
Start time(s)***
Wait time(s)5
Collimation0.5 x 8
Pitch***
kV100
mA80
Rotation time(s)0.275
Total scan time(s)40.425
Range(mm)0
Direction***
CFOVM
CEON
CTDI(mGy)340
DLP(mGy.cm)136
ReconstructionAxial1
Slice thickness4.0 mm
Slice Interval***
DFOV220
SUREIQCardiac - Low Dose
Recon Process***
FC2
Boost 3DOFF
OSR***
FilterOFF
Volume - 1cm above LM coronary to 1cm below apex of heart
Scan modeVolume ECG Gated Modulated
# of scan(s)1
Start time(s)0
Wait time(s)0
Collimation0.5
Pitch***
kV100
mAExp.3D CTA Standard
Rotation time(s)0.275
Total scan time(s)1.176
Range(mm)120
DirectionOUT
CFOVM
CEON
CTDI(mGy)***
DLP(mGy.cm)***

 

ReconstructionVolume1 (75%)Volume2 (PhaseXact)Volume3 (0 - 90% R-R every 10%)
Slice thickness0.5 mm0.5 mm1.0 mm
Slice Interval0.50 mm0.50 mm1.00 mm
DFOV220220220
SUREIQCardiac - CTA StandCardiac - CTA StandCardiac - CTA Stand
Recon ProcessAIDR 3D StandardAIDR 3D StandardAIDR 3D Standard
FC333
Boost 3DOFFOFFOFF
OSRCardiacCardiacCardiac
FilterOFFOFFOFF

 

* Exp.3D Adult ECG CTA Standard Preset - Reference Image Quality
Target SD33
Target Slice thickness0.5 mm
Target SUREIQCardiac CTA Stand
Recon ProcessAIDR 3D Standard
FC3
XY modulation***
Max(mA)900
Min(mA)40