EP No: 102

Scanogram - AP Scanogram Above shoulders to below liver
Scan modeScanogram(AP)
Start time(s)0
Wait time(s)0
kV120
mA50
Range(mm)500
DirectionOUT
Display filterSharp (FL03)
Scano angle °0
CEOFF

 

Scanogram - Lateral Scanogram
Scan modeScanogram(Lat)
Start time(s)0
Wait time(s)6.1
kV120
mA100
Range(mm)500
DirectionIN
Display filterSharp (FL03)
Scano angle °90
CEOFF

 

Volume - Carina to below apex of heart
Scan modeVolume Ca Score
# of scan(s)1
Start time(s)0
Wait time(s)0
Collimation0.5
Pitch***
kV120
mAExp.3D Ca Score
Rotation time(s)0.275
Total scan time(s)0.275
Range(mm)140
DirectionOUT
CFOVM
CEOFF
CTDI(mGy)***
DLP(mGy.cm)***

 

ReconstructionVolume1MutiView1 Axial
Slice thickness0.5 mm3.0 mm
Slice Interval0.50 mm3.0 mm
DFOV220***
SUREIQCardiac - Ca ScoreCardiac - Ca Score
Recon ProcessOFF***
FC12***
Boost 3DOFF***
OSRCardiac***
FilterOFF***

 

* Exp.3D Adult ECG Ca Score Preset - Reference Image Quality
Target SD55
Target Slice thickness0.5 mm
Target SUREIQCardiac Ca Score
Recon ProcessOFF
FC12
XY modulation***
Max(mA)300
Min(mA)40