Specific Anatomic RegionLiver
ApplicationLiving related liver donor
AuthorElliot K. Fishman, MD
Reference SourcePersonal communication
Scanner UsedSensation 16
KV / Effective mAs / Rotation time (sec)120 kV/ 225 eff. mAs / 0.5 sec
Detector Collimation (mm)  0.75 mm
Slice Thickness (mm)  0.75 mm
Feed / Rotation (mm)  12.0 mm
Kernel  B30f medium smooth
Increment (mm)0.75 mm x 0.75 mm (3D) 5 mm x 5mm (filming)
Image Ordercraniocaudal
Oral Contrastwater (1 12 oz. cup = 1000 cc)
IV Contrast Volume and Type120 cc of Omnipaque 350
Injection Rate3.0-3.5 cc/sec
Scan Delay (sec)Arterial: 25-30 sec Venous: 20 sec
3D Technique UsedVRT and MIP  
Comment: -We also use the 5mm x 5mm sections to do liver volumes.