Acquisition Phase | Phase Included | Scan Delay | Respiration Phase | Anatomical Coverage |
Non Contrast | N/A | N/A | N/A | N/A |
Arterial Phase | Yes | Bolus triggered | Inspiration | Lung apices through the diaphragm |
Venous Phase | N/A | N/A | N/A | N/A |
Delayed Phase | N/A | N/A | N/A | N/A |
Scan Comments: The flash spiral mode is most successful when imaging patients with heart rates 60-70 bpm. Utilizing this mode will result in extremely low dose exams. All scans are performed using the ALARA principle. 3D post-processing is performed with this protocol. |
Technical Parameters | Parameters |
kVp | 120 |
Effective mAs | Use dose reduction software |
Quality Reference mAs | 360 |
Time (Rotation) | 0.28 sec |
Average Acquisition Time | 1-2 sec |
Collimation | 128 x 0.6mm |
Pitch Value | 3.2 |
Scan Direction | Craniocaudal |
Comments: Using the flash spiral mode allows for greatest dose savings potential. The entire data series is collected in one heart beat thus avoiding the need for a slow pitch with multiple images across the R-R interval of the ECG wave. 100 kVp is used for most average size patients. ECG dose modulation is not necessary for this protocol. Dose reduction software is used to reduce patient dose. |
Reconstruction Parameters | Thin Data Heart Mag | Soft Tissue Open Field | Thin Data Open Field | Lung Open Field |
Slice Thickness | 0.75 mm | 3 mm | 0.75 mm | 3 mm |
Reconstruction Spacing | 0.5 mm | 3 mm | 0.5 mm | 3 mm |
Reconstruction Algorithm | B26f ASA | B30f | B20f | B80f |
Window Width and Level | 500/100 | 410/10 | 410/10 | 1600/-500 |
Reconstruction Comments: The reconstruction field of view is limited to 33.2cm using the flash spiral scan mode. An additional smaller DFOV reconstruction that only includes the heart and aorta will increase spatial resolution of the coronary vessels. |
Reconstruction Parameters | Coronal Open Field | Sagittal Open Field |
Slice Thickness | 3 mm | 3 mm |
Reconstruction Spacing | 3 mm | 3 mm |
Reconstruction Algorithm | B30f | B30f |
Window Width and Level | 410/10 | 410/10 |
Reconstruction Comments: The reconstruction field of view is limited to 33.2cm using the flash spiral scan mode. An additional smaller DFOV reconstruction that only includes the heart and aorta will increase spatial resolution of the coronary vessels. |
Contrast Parameters | Parameters |
Contrast Type | Non-ionic |
Contrast Volume | 80-100 mL |
Saline Flush | 30 mL |
Injection Rate | 5-7 mL/sec |
Oral Contrast | N/A |
Contrast Volume | N/A |
Comments: Bolus tracking is used ideal with a 300 HU trigger value on the ascending aorta. A saline flush can be used to soften the SVC and limit beam hardening artifact. |
Other Comments: To successfully scan a patient using the flash spiral mode, the heart rate should be 60-70bpm and the patient should be no larger than average size to avoid increased image noise. Typically, flash spiral gated chest acquisitions are performed with an ECG trigger of 60%. It may also be useful to trigger at 30% for patients with higher heart rates. The ECG trigger must be changed utilizing the scan protocol assistant, so it may be beneficial to build two separate protocols, diastolic and systolic, when programming the scanner.