EP No: 101
Scanogram - AP Scanogram Above shoulders to below liver | |
Scan mode | Scanogram(AP) |
Start time(s) | 0 |
Wait time(s) | 0 |
kV | 120 |
mA | 50 |
Range(mm) | 500 |
Direction | OUT |
Display filter | Sharp (FL03) |
Scano angle ° | 0 |
CE | OFF |
Scanogram - Lateral Scanogram | |
Scan mode | Scanogram(Lat) |
Start time(s) | 0 |
Wait time(s) | 6.1 |
kV | 120 |
mA | 100 |
Range(mm) | 500 |
Direction | IN |
Display filter | Sharp (FL03) |
Scano angle ° | 90 |
CE | OFF |
S&V - Position at aortic arch Trigger at 180 HU | |
Scan mode | S&V |
# of scan(s) | 1 |
Start time(s) | *** |
Wait time(s) | 0 |
Collimation | 1.0 x 4 |
Pitch | *** |
kV | 100 |
mA | 80 |
Rotation time(s) | 0.275 |
Total scan time(s) | 0.275 |
Range(mm) | 4 |
Direction | OUT |
CFOV | L |
CE | OFF |
CTDI(mGy) | 2.7 |
DLP(mGy.cm) | 1.1 |
Reconstruction | Axial1 |
Slice thickness | 4.0 mm |
Slice Interval | *** |
DFOV | 350 |
SUREIQ | Body - Low Dose |
Recon Process | OFF |
FC | 2 |
Boost 3D | OFF |
OSR | *** |
Filter | OFF |
SureStart - Trigger in aortic arch at 180 HU | |
Scan mode | SureStart |
# of scan(s) | *** |
Start time(s) | *** |
Wait time(s) | 10 |
Collimation | 0.5 x 8 |
Pitch | *** |
kV | 100 |
mA | 80 |
Rotation time(s) | 0.275 |
Total scan time(s) | 40.15 |
Range(mm) | 0 |
Direction | *** |
CFOV | L |
CE | ON |
CTDI(mGy) | 375.7 |
DLP(mGy.cm) | 150.3 |
Reconstruction | Axial1 |
Slice thickness | 4.0 mm |
Slice Interval | *** |
DFOV | 350 |
SUREIQ | Body - Low Dose |
Recon Process | *** |
FC | 2 |
Boost 3D | OFF |
OSR | *** |
Filter | OFF |
Helical - Above aortic arch to below lung bases | |
Scan mode | Helical |
# of scan(s) | 1 |
Start time(s) | 0 |
Wait time(s) | 0 |
Collimation | 0.5 x 80 |
Pitch | Standard |
kV | 100 |
mA | Exp.3D Standard |
Rotation time(s) | 0.275 |
Total scan time(s) | 3.1 |
Range(mm) | 300 |
Direction | OUT |
CFOV | L |
CE | ON |
CTDI(mGy) | *** |
DLP(mGy.cm) | *** |
Reconstruction | Axial1 | Volume1 |
Slice thickness | 3.0 mm | 0.5 mm |
Slice Interval | 3.00 mm | 0.30 mm |
DFOV | 350 | 350 |
SUREIQ | CTA - Body | CTA - Body |
Recon Process | AIDR 3D Standard | AIDR 3D Standard |
FC | 8 | 8 |
Boost 3D | OFF | OFF |
OSR | ON | ON |
Filter | OFF | OFF |
* Exp.3D Adult Chest Standard Preset - Reference Image Quality | |
Target SD | 12.5 |
Target Slice thickness | 5.0 mm |
Target SUREIQ | Body Std. Axial |
Recon Process | AIDR 3D Standard |
FC | 18 |
XY modulation | ON |
Max(mA) | 700 |
Min(mA) | 80 |