EP No: 99

Scanogram - AP Scanogram Above shoulders to below liver
Scan modeScanogram(AP)
Start time(s)0
Wait time(s)0
kV120
mA50
Range(mm)500
DirectionOUT
Display filterSharp (FL03)
Scano angle °0
CEOFF

 

Scanogram - Lateral Scanogram
Scan modeScanogram(Lat)
Start time(s)0
Wait time(s)6.1
kV120
mA100
Range(mm)500
DirectionIN
Display filterSharp (FL03)
Scano angle °90
CEOFF

 

Helical - Above lung apices to below lung bases
Scan modeHelical
# of scan(s)1
Start time(s)0
Wait time(s)0
Collimation0.5 x 80
PitchStandard
kV120
mAExp.3D Standard
Rotation time(s)0.275
Total scan time(s)3.6
Range(mm)360
DirectionOUT
CFOVL
CEOFF
CTDI(mGy)***
DLP(mGy.cm)***


AIDR 3D Standard

ReconstructionAxial1Axial3Volume1MutiView1 CoronalMutiView1 Sagittal
Slice thickness1.0 mm5.0 mm0.5 mm3.0 mm3.0 mm
Slice Interval10.00 mm5.00 mm0.30 mm3.0 mm3.0 mm
DFOV400400400******
SUREIQLung - Std. AxialBody - Std. AxialLung - Std. VolumeLung - Std. VolumeLung - Std. Volume
Recon ProcessAIDR 3D StandardAIDR 3D StandardAIDR 3D Standard******
FC521851******
Boost 3DOFFOFFOFF******
OSRONONON******
FilterOFFOFFOFF******

 

Exp.3D Adult Chest Standard Preset - Reference Image Quality
Target SD12.5
Target Slice thickness5.0 mm
Target SUREIQBody Std. Axial
Recon ProcessAIDR 3D Standard
FC18
XY modulationON
Max(mA)700
Min(mA)80