EP No: 106

Scanogram - Lateral Scanogram Below base of skull to vertex
Scan modeScanogram(AP)
Start time(s)0
Wait time(s)0
kV120
mA50
Range(mm)500
DirectionOUT
Display filterSharp (FL03)
Scano angle °0
CEOFF

 

Scanogram - Lateral Scanogram
Scan modeScanogram(Lat)
Start time(s)0
Wait time(s)6.1
kV120
mA100
Range(mm)500
DirectionIN
Display filterSharp (FL03)
Scano angle °90
CEOFF

 

S&V - Position in aortic arch
Scan modeS&V
# of scan(s)1
Start time(s)***
Wait time(s)0
Collimation1.0 x 4
Pitch***
kV0
mA80
Rotation time(s)0.275
Total scan time(s)0.275
Range(mm)4
DirectionOUT
CFOVM
CEOFF
CTDI(mGy)2.4
DLP(mGy.cm)1

 

ReconstructionAxial1
Slice thickness4.0 mm
Slice Interval***
DFOV320
SUREIQCardiac - Low Dose
Recon ProcessOFF
FC2
Boost 3DOFF
OSR***
FilterOFF
SureStart - Trigger in aorta at 180 HU
Scan modeSureStart
# of scan(s)***
Start time(s)***
Wait time(s)10
Collimation0.5 x 8
Pitch***
kV100
mA80
Rotation time(s)0.275
Total scan time(s)40.425
Range(mm)0
Direction***
CFOVM
CEON
CTDI(mGy)340
DLP(mGy.cm)136
ReconstructionAxial1
Slice thickness4.0 mm
Slice Interval***
DFOV320
SUREIQCardiac - Low Dose
Recon Process***
FC2
Boost 3DOFF
OSR***
FilterOFF

 

Helical - From subclavian artery to below apex of heart
Scan modeHelical ECG Gated Prospective
# of scan(s)1
Start time(s)0
Wait time(s)0
Collimation0.5 x 160
PitchStandard
kV100
mAExp.3D CTA Standard
Rotation time(s)0.275
Total scan time(s)6.9
Range(mm)321
DirectionOUT
CFOVM
CEON
CTDI(mGy)***
DLP(mGy.cm)***

 

ReconstructionVolume1 (75%)Volume2 (PhaseXact)Volume3 (75%)MutiView1 AxialMutiView3 Axial
Slice thickness0.5 mm0.5 mm0.5 mm5.0 mm5.0 mm
Slice Interval0.30 mm0.30 mm0.30 mm5.0 mm5.0 mm
DFOV320320320******
SUREIQCardiac - CTA StandCardiac - CTA StandLung - Std. VolumeCardiac - CTA StandLung - Std. Volume
Recon ProcessAIDR 3D StandardAIDR 3D StandardAIDR 3D Standard******
FC3351******
Boost 3DOFFOFFOFF******
OSRONONON******
FilterOFFOFFOFF******

 

* Exp.3D Adult Neck Quality Preset - Reference Image Quality
Target SD33
Target Slice thickness0.5 mm
Target SUREIQCardiac CTA Stand
Recon ProcessAIDR 3D Standard
FC3
XY modulation***
Max(mA)900
Min(mA)40