EP No: 108

Scanogram - AP Scanogram Above shoulders to below liver
Scan modeScanogram(AP)
Start time(s)0
Wait time(s)0
kV120
mA50
Range(mm)500
DirectionOUT
Display filterSharp (FL03)
Scano angle °0
CEOFF

 

Scanogram - Lateral Scanogram
Scan modeScanogram(Lat)
Start time(s)0
Wait time(s)0
kV120
mA100
Range(mm)500
DirectionIN
Display filterSharp (FL03)
Scano angle °90
CEOFF

 

Dy-Volume - Position over area of interest
Scan modeDy-Volume Continuous
# of scan(s)C1
Start time(s)0
Wait time(s)0
Collimation0.5
Pitch***
kV120
mA50
Rotation time(s)0.5
Total scan time(s)10.5
Range(mm)160
DirectionOUT
CFOVL
CEOFF
CTDI(mGy)54.3
DLP(mGy.cm)868.8
Continuous Setting Start Time 0.0 s , End Time 10.0 s

 

ReconstructionVolume1
Slice thickness0.5 mm
Slice Interval0.50 mm
DFOV400
SUREIQBody - Std. Volume
Recon ProcessAIDR 3D Standard
FC8
Boost 3DOFF
OSRBody
FilterOFF