History:Airways disease. Check with radiologist to see if HI-res is necessary.
Surview:Sternal notch to mid kidneys

 

Scan:Two scans: apices to diaphragm.
Scan #1:BREATHING INSTRUCTIONS: Inspiration
Scan #2 :BREATHING INSTRUCTIONS: Expiration - 3 levels, 2 contiguous axial slices:
1. level of arch
2. level of carina
3. level of bases

 

Thick/Couch Incr.:
4 @ 31.25 @ 1.2
Length
 2.45
Kvp
120120
mAs
250250
Resolution
standardHigh
Collimation
128 x 0.6252 x 0.625
Pitch
0.993N/A
Rotation Time
0.50.33
Scan FOV
350350
Cycle TimeN/A0.9
CyclesN/A2
Filter
BL
iDose4
iD-3iD-3
Lung reconL/-400/1600-400/1600
Recon
0.9 @ 0.45-B filter
Matrix
512768

 

**Note: There will be three levels for scans. See levels listed above. DUPLICATE EACH SCAN and move to the correct level. You will have 2 contiguous images for each level.

DO NOT SCAN TO THE LEVEL OF THE ADRENALS

DOSE Tools: Z-Dom

Post Processing

MPRCoronal recon using 0.9 @ 0.45 dataRecon
4 @ 3

 Archiving

Lung windows
4 @ 3
Axial-Hi Res
1.25 @ 1.2
Mediastinal windows
0.9 @ 0.45
MPR
4 @ 3

 

Courtesy of the University of Maryland