Acquisition Phase | Phase Included | Scan Delay | Respiration Phase | Anatomical Coverage |
Non Contrast | N/A | N/A | N/A | N/A |
Arterial Phase | N/A | N/A | N/A | N/A |
Venous Phase | Yes | Patient Dependent | Inspiration | Region of interest |
Delayed Phase | N/A | N/A | N/A | N/A |
Scan Comments: All scans are performed using the ALARA Principle. | ||||
Technical Parameters | Parameters |
kVp | 120 |
Effective mAs | See chart |
Care Dose Reference mAs | N/A |
Time (Rotation) | 1.0 |
Average Acquisition Time | 6-8 sec |
Collimation | 64 x 0.6mm |
Pitch Value | 0.8 |
Scan Direction | Craniocaudal |
Comments: The effective mAs is adjusted using guidelines based upon the patient’s weight. | |
Reconstruction Parameters | Soft Tissue | Thin Data | Bone |
Slice Thickness | 3mm | 0.75mm | 3mm |
Reconstruction Spacing | 3mm | 0.5mm | 3mm |
Reconstruction Algorithm | B30s | B60s | B70s |
Window Width and Level | 350 / 40 | 2500 / 500 | 2500 / 500 |
Reconstruction Comments: Coronal and sagittal reconstructions are performed. True axial & coronal MPR images are generated when patients are unable to be positioned correctly. | |||
Contrast Parameters | Parameters |
Contrast Type | N/A |
Contrast Volume | N/A |
Saline Flush | N/A |
Injection Rate | N/A |
Oral Contrast | N/A |
Contrast Volume | N/A |
Comments: IV contrast is typically not necessary. | |
The effective mAs for a pediatric spine CT is calculated using the following chart.
Weight (kg) | Effective mAs |
<15 | 50-70 mAs |
15-24 | 65-80 mAs |
25-34 | 80-95 mAs |
35-44 | 95-115 mAs |
45< | 115-150mAs |
Acquisition Phase | Phase Included | Scan Delay | Respiration Phase | Anatomical Coverage |
Non Contrast | N/A | N/A | N/A | N/A |
Arterial Phase | N/A | N/A | N/A | N/A |
Venous Phase | Yes | Patient Dependent | Inspiration | Region of interest |
Delayed Phase | N/A | N/A | N/A | N/A |
Scan Comments: All scans are performed using the ALARA Principle. | ||||
Technical Parameters | Parameters |
kVp | 120 |
Effective mAs | See chart |
Care Dose Reference mAs | N/A |
Time (Rotation) | 1.0 |
Average Acquisition Time | 6-8 sec |
Collimation | 64 x 0.6mm |
Pitch Value | 0.8 |
Scan Direction | Craniocaudal |
Comments: The effective mAs is adjusted using guidelines based upon the patient’s weight. | |
Reconstruction Parameters | Soft Tissue | Thin Data | Bone |
Slice Thickness | 3mm | 0.75mm | 3mm |
Reconstruction Spacing | 3mm | 0.5mm | 3mm |
Reconstruction Algorithm | B30s | B60s | B70s |
Window Width and Level | 350 / 40 | 2500 / 500 | 2500 / 500 |
Reconstruction Comments: Coronal and sagittal reconstructions are performed. True axial & coronal MPR images are generated when patients are unable to be positioned correctly. | |||
Contrast Parameters | Parameters |
Contrast Type | N/A |
Contrast Volume | N/A |
Saline Flush | N/A |
Injection Rate | N/A |
Oral Contrast | N/A |
Contrast Volume | N/A |
Comments: IV contrast is typically not necessary. | |
The effective mAs for a pediatric spine CT is calculated using the following chart.
Weight (kg) | Effective mAs |
<15 | 50-70 mAs |
15-24 | 65-80 mAs |
25-34 | 80-95 mAs |
35-44 | 95-115 mAs |
45< | 115-150mAs |