Acquisition Phase | Phase Included | Scan Delay | Respiration Phase | Anatomical Coverage |
Non Contrast | Yes | N/A | Inspiration | Aortic stent only |
Arterial Phase | Yes | Bolus triggered | Inspiration | Entire aortic vasculature |
Venous Phase | N/A | N/A | N/A | N/A |
Delayed Phase | Yes | 60 sec post-arterial phase imaging | Inspiration | Aortic stent only |
Scan Comments: Non-contrast and delayed images may be necessary when following up stent placement or to r/o endovascular leak. Alternatively, dual energy can be used to provide virtual non-contrast images. All scans are performed using the ALARA principle. 3D post-processing is performed with this protocol. |
Technical Parameters | Parameters |
kVp | 120 |
Effective mAs | Use dose reduction software |
Quality Reference mAs | 290 |
Time (Rotation) | 0.5 sec |
Average Acquisition Time | 5-7 sec |
Collimation | 128 x 0.6mm |
Pitch Value | 0.6 |
Scan Direction | Craniocaudal |
Comments: Dose reduction software is used to reduce patient dose. The quality reference mAs can be lowered on non-contrast and delayed images (250 mAs). |
Reconstruction Parameters | Soft Tissue | Thin Data | Thin Data | Lung |
Slice Thickness | 3 mm | 0.75 mm | 0.75 mm | 3 mm |
Reconstruction Spacing | 3 mm | 0.5 mm | 0.5 mm | 3 mm |
Reconstruction Algorithm | I40f (2) | I26f (2) | B20f | I70f (2) |
Window Width and Level | 410/10 | 410/10 | 410/10 | 1600/-500 |
Reconstruction Comments: |
Reconstruction Parameters | Coronal | Sagittal |
Slice Thickness | 3 mm | 3 mm |
Reconstruction Spacing | 3 mm | 3 mm |
Reconstruction Algorithm | I40f (2) | I40f (2) |
Window Width and Level | 410/10 | 410/10 |
Reconstruction Comments: |
Contrast Parameters | Parameters |
Contrast Type | Non-ionic |
Contrast Volume | 120 mL |
Saline Flush | N/A |
Injection Rate | 4-5 mL/sec |
Oral Contrast | H2O |
Contrast Volume | 1000 mL |
Comments: Bolus tracking is performed off the abdominal aorta using a 230 HU trigger setting. |