Acquisition Phase | Phase Included | Scan Delay | Respiration Phase | Anatomical Coverage |
Non Contrast | N/A | N/A | N/A | N/A |
Arterial Phase | Yes | Bolus triggered | Inspiration | diaphragm through pubic symphysis |
Venous Phase | Yes | 60 sec | Inspiration | diaphragm through iliac crest |
Delayed Phase | Yes | 5-7 min | Inspiration | kidneys through bladder |
Scan Comments: Scan ranges are limited to a specific area of interest to avoid unnecessary patient dose. Dual energy post-processing software can be used to eliminate the need for acquiring non-contrast images. Dual energy is used only for the arterial phase images in order to reduce the number of images generated for each study. Single energy is used for all other phases. For specific technical parameters, see the single energy kidney mass protocol. All scans are performed using the ALARA principle. 3D post-processing is performed with this protocol. |
Technical Parameters | Parameters |
kVp | 100/140 |
Effective mAs | Use dose reduction software |
Quality Reference mAs | 250/193 |
Time (Rotation) | 0.5 sec |
Average Acquisition Time | 10-15 sec |
Collimation | 32 x 0.6mm |
Pitch Value | 0.8 |
Scan Direction | Craniocaudal |
Comments: Images are acquired using 100/140 kvp setting with a Sn filter. Dose reduction software is used to reduce patient dose. The pitch value should remain below 1.0 to reduce any artifacts during post-processing. |
Reconstruction Parameters | Soft Tissue | Thin Data | Lung | Coronal |
Slice Thickness | 3 mm | 0.75 mm | 3 mm | 3 mm |
Reconstruction Spacing | 3 mm | 0.5 mm | 3 mm | 3 mm |
Reconstruction Algorithm | B31f | B20f | B80f | B31f |
Window Width and Level | 410/10 | 410/10 | 1600/-500 | 410/10 |
Reconstruction Comments: |
Reconstruction Parameters | Sagittal |
Slice Thickness | 3 mm |
Reconstruction Spacing | 3 mm |
Reconstruction Algorithm | B31f |
Window Width and Level | 410/10 |
Reconstruction Comments: |
Contrast Parameters | Parameters |
Contrast Type | Non-ionic |
Contrast Volume | 120 mL |
Saline Flush | N/A |
Injection Rate | 4-5 mL/sec |
Oral Contrast | H2O |
Contrast Volume | 1000 mL |
Comments: Bolus tracking is performed off the abdominal aorta using a 230 HU trigger setting. |