Acquisition Phase | Phase Included | Scan Delay | Respiration Phase | Anatomical Coverage |
Non Contrast | N/A | N/A | N/A | N/A |
Arterial Phase | Yes | Bolus triggered | Inspiration | Lung apices through the diaphragm to include the adrenal glands |
Venous Phase | N/A | N/A | N/A | N/A |
Delayed Phase | N/A | N/A | N/A | N/A |
Scan Comments: A flash spiral mode can be used to improve scan speed and limit breathing artifact if the patient is short of breath. All scans are performed using ALARA principle. |
Technical Parameters | Parameters |
kVp | Care kV (ref. kV 120) |
Effective mAs | Use dose reduction software |
Care Dose Reference mAs | 170 |
Time (Rotation) | 0.5 sec |
Average Acquisition Time | 4-6 sec |
Collimation | 128 x 0.6mm |
Pitch Value | 1.0 |
Scan Direction | Craniocaudal |
Comments: Dose reduction software is used to reduce patient dose. 100 kVp can be used for most average size patients which will significantly reduce patient dose. |
Reconstruction Parameters | Soft Tissue | Thin Data | Lung | Coronal |
Slice Thickness | 3 mm | 0.75 mm | 3 mm | 2 mm |
Reconstruction Spacing | 3 mm | 0.5 mm | 3 mm | 2 mm |
Reconstruction Algorithm | B30f | B20f | B80f | B30f |
Window Width and Level | 550/50 | 550/50 | 1600/-500 | 410/10 |
Reconstruction Comments: |
Reconstruction Parameters | Sagittal | Sliding MIP Axial | |
Slice Thickness | 2 mm | 16 mm | |
Reconstruction Spacing | 2 mm | 3 mm | |
Reconstruction Algorithm | B30f | B50f | |
Window Width and Level | 410/10 | 700/80 | |
Reconstruction Comments: |
Contrast Parameters | Parameters |
Contrast Type | Non-ionic |
Contrast Volume | 80-120mL |
Saline Flush | N/A |
Injection Rate | 5-6 mL/sec |
Oral Contrast | N/A |
Contrast Volume | N/A |
Comments: Bolus tracking is performed off the pulmonary trunk using a 200 HU trigger setting. |