Acquisition Phase | Phase Included | Scan Delay | Respiration Phase | Anatomical Coverage |
Non Contrast | Yes | N/A | Inspiration | Nasopharynx through the diaphragm |
Arterial Phase | N/A | N/A | N/A | N/A |
Venous Phase | N/A | N/A | N/A | N/A |
Delayed Phase | N/A | N/A | N/A | N/A |
Scan Comments: A high resolution flash spiral scan may be useful if the patient is short of breath to avoid breathing artifact. All scans are performed using ALARA principle. 3D post-processing is performed with this protocol. |
Technical Parameters | Parameters |
kVp | Care kV (ref. kV 120) |
Effective mAs | Use dose reduction software |
Care Dose Reference mAs | 185 |
Time (Rotation) | 0.5 sec |
Average Acquisition Time | 4-6 sec |
Collimation | 128 x 0.6mm |
Pitch Value | 1.0 |
Scan Direction | Craniocaudal |
Comments: Dose reduction software is used to reduce patient dose and reduce image noise, specifically in the shoulder region. 100 kVp can also be used on average size patients. |
Reconstruction Parameters | Soft Tissue | Thin Data | Lung | Coronal |
Slice Thickness | 3 mm | 0.75 mm | 3 mm | 3 mm |
Reconstruction Spacing | 3 mm | 0.5 mm | 3 mm | 3 mm |
Reconstruction Algorithm | B30f | B20f | B80f | B20f |
Window Width and Level | 410/10 | 410/10 | 1600/-500 | 410/10 |
Reconstruction Comments: Coronal and sagittal image reconstructions are performed aligned to the patient’s trachea. |
Reconstruction Parameters | Sagittal | ||
Slice Thickness | 3 mm | ||
Reconstruction Spacing | 3 mm | ||
Reconstruction Algorithm | B20f | ||
Window Width and Level | 410/10 | ||
Reconstruction Comments: Coronal and sagittal image reconstructions are performed aligned to the patient’s trachea. |
Contrast Parameters | Parameters |
Contrast Type | N/A |
Contrast Volume | N/A |
Saline Flush | N/A |
Injection Rate | N/A |
Oral Contrast | N/A |
Contrast Volume | N/A |
Comments: IV contrast is not typically necessary for this exam. |