| kV | 80 | ||
| Effective mAs | * | ||
| Slice collimation | 0.75 mm | ||
| Slice width | 0.75 mm | 2nd reconstruction 3.0 mm | 3rd reconstruction 3.0 mm |
| Feed / Rotation | 18.0 mm | 18.0 mm | |
| Rotation time | 0.5 sec | 0.5 sec | |
| Kernel | B30f | B30f | B80f |
| Increment | 0.5 mm | 3.0 mm | 3.0 mm |
| Image Order | cr-ca | Lung | |
| Comment: The above is for the arterial phase. You will also scan a venous phase using the same recon jobs except recon #3 will be B30f kernal using a liver window. * The mAs should be adjusted to the body weight of the child: 1. < 15 kg: kV:80, mAs: 70 2. 15-24 kg : kV: 80, mAs: 80 3. 25-34 kg : kV: 80, mAs: 100 4. 35-44 kg : kV: 80, mAs: 120 5. 45-54 kg : kV: 80, mAs: 180 | |||