| Specific Anatomic Region | Abdomen |
| Application | Trauma |
| Author | Shuman WP |
| Reference Source | Radiology 1997; 205:297-306 |
| Scanner Used | Not defined |
| Injection Rate | 2-4ml/sec |
| Contrast Volume and Type | 130-180ml of a 60-75% reconstruction of non-ionic contrast |
| Area Scanned | Inferior lungs to inferior edge of ischia |
| Scan Delay | 70-90sec |
| Length of Spiral (time) | As needed (~ 40sec) |
| Slice Thickness | 5-8mm |
| Table Speed/Pitch | Pitch of 1-2 |
| Reconstruction Interval | 5-8mm |
| 3D Technique Used | None |
Comment: - This is a general exam. Focused exams may require arterial or delayed phase imaging.
- 3D reconstruction may be needed depending on the extent of trauma.
|