| Specific Anatomic Region | Thoracic Aorta |
| Application | R/O dissection |
| Author | Fishman EK |
| Reference Source | Personal communication |
| Scanner Used | Siemens Somatom Plus4 |
| Injection Rate | 3ml/sec |
| Contrast Volume and Type | 100-120ml of Omnipaque-350 |
| Area Scanned | From diaphragm to lung apex |
| Scan Delay | 30sec |
| Length of Spiral (time) | 30-40sec |
| Slice Thickness | 3mm |
| Table Speed/Pitch | 5-6mm/sec |
| Reconstruction Interval | 2-3mm |
| 3D Technique Used | Volume rendering |
| Comment: We scan from the bases upward to decrease the artifact off contrast in the axilla as well as to take advantage of optimizing patient breathhold. |