| Specific Anatomic Region | Chest |
| Application | Stage lymphoma |
| Author | Fishman EK |
| Reference Source | RadioGraphics 1991; 11(4): 647-669 |
| Scanner Used | Siemens Somatom Plus 4 |
| Injection Rate | 2cc/sec |
| Contrast Volume and Type | 100cc of Omnipaque 350 |
| Area Scanned | Apex of lung through bases |
| Scan Delay | 40sec |
| Length of Spiral (time) | 25-35sec |
| Slice Thickness | 5mm |
| Table Speed/Pitch | 8mm/sec or a pitch of 1.6 |
| Reconstruction Interval | 5mm |
| 3D Technique Used | None usually needed |
| Comment: Scans may begin in the neck using a 5x5x5 protocol (5mm collimation, 5mm/sec table speed, and 5mm interscan reconstruction) if desired. |