| Specific Anatomic Region | Liver | |
| Application | Suspected metastases from colon cancer | |
| Author | Fishman EK | |
| Reference Source | Personal Communication | |
| Scanner Used | Siemens Plus4 Volume Zoom | |
| KV / mAs / Time per Rotation (sec) | 120/165/.5 | |
| Detector Collimation (mm) | 2.5 | |
| Slice Thickness (mm) | 5 | |
| Data Reconstruction Interval (mm) | 5 | |
| Table Speed (mm per rotation) / Pitch | 15 / 6 | |
| Oral Contrast | 750 cc of 3% Hypaque | |
| IV Contrast Volume and Type | 120 cc of Omnipaque 350 | |
| Injection Rate | 2-3 cc/sec | |
| Scan Delay (sec) | 50-60 sec | |
| 3D Technique Used | N/A | |
| Comment: This is a single phase protocol. For evaluation of a tumor like colon cancer, which has hypovascular metastases, a "portal phase only study" is needed. Arterial phase imaging is of no additional value. | ||