| Specific Anatomic Region | Liver | |
| Application | Differential diagnosis of a liver mass | |
| Author | Fishman EK | |
| Reference Source | Personal Communication | |
| Scanner Used | Siemens Plus4 Volume Zoom | |
| KV / mAs / Time per Rotation (sec) | 120/165/.5 | |
| Detector Collimation (mm) | 2.5 | |
| Slice Thickness (mm) | 3 | |
| Data Reconstruction Interval (mm) | 3-5 | |
| Table Speed (mm per rotation) / Pitch | 12.5/5 | |
| Oral Contrast | 750 cc of water | |
| IV Contrast Volume and Type | 120 cc of Omnipaque 350 | |
| Injection Rate | 3 cc/sec | |
| Scan Delay (sec) | 30 sec arterial & then delayed scans | |
| 3D Technique Used | Volume rendering in select cases | |
| Comments: This will require multiple phases of data acquisition which always include arterial and portal phase imaging. Additional delayed phase imaging may be helpful in cases of suspected cholangiocarcinoma. | ||