Specific Anatomic Region | Chest | |
Application | Suspected pulmonary embolism (classic) | |
Author | Fishman EK | |
Reference Source | Personal Communication | |
Scanner Used | Siemens Plus4 Volume Zoom | |
KV / mAs / Time per Rotation (sec) | 140/100/.5 | |
Detector Collimation (mm) | 2.5 | |
Slice Thickness (mm) | 3 | |
Data Reconstruction Interval (mm) | 2 | |
Table Speed (mm per rotation) / Pitch | 15/6 | |
Oral Contrast | N/A | |
IV Contrast Volume and Type | 120 cc of Omnipaque 350 | |
Injection Rate | 3 cc/sec | |
Scan Delay (sec) | 30 sec | |
3D Technique Used | MPR or volume rendering in select cases | |
Comment: This matches the classic technique used with single detector CT. A combination of 1 mm slice thickness of collimation may be considered for PE & is under investigation. |