Siemens Definition 64 Slice MDCT Protocols : Cardiac : DS_CoronaryCTAVol (Adult) | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
Source: Scanner Default | ||||||||||
| Protocol | Range | Series Description | kV | Quality Ref. mAs | (Eff.) mAs | CAREDose/CAREDose 4D | CAREDoseType | CTDIvol | RotTime | Pitch |
DS_CoronaryCTAVol (Adult) | Topogram | Topogram 0.6 T20s | 120 | 0 | 186 | off | CAREDose | 0 | 0.500 | 1.0000000 |
| DS_CorCTA |
| 120 | 320 | 320 | on | CAREDose4D | 45.9344 | 0.330 | 0.20 |
|
| DS_CorCTA 3.0 B30f 65% |
|
|
|
|
|
|
|
|
|
| DS_CorCTA 0.6 B26f 65% |
|
|
|
|
|
|
|
|
|
| RAO DS_CorCTA 2.0 MIP 65% |
|
|
|
|
|
|
|
|
|
| LAO DS_CorCTA 2.0 MIP 65% |
|
|
|
|
|
|
|
|
|
| CX DS_CorCTA 2.0 MIP 65% |
|
|
|
|
|
|
|
|
| Coll. | Slice | Recon increment | Images | Kernel | Window | API | Comment1 |
|---|---|---|---|---|---|---|---|
0.60 | 0.60 |
| 1 | T20s | Topogram Body | None |
|
0.60 |
|
|
|
|
| None |
|
| 3.00 | 3.0 | 41 | B30f | Cardiac |
|
|
| 0.60 | 0.3 | 401 | B26f | Cardiac |
|
|
| 2.00 | 2.0 | 62 | B30f | Cardiac |
|
|
| 2.00 | 2.0 | 73 | B30f | Cardiac |
|
|
| 2.00 | 2.0 | 45 | B30f | Cardiac |
|
|