Siemens Definition 64 Slice MDCT Protocols : Chest : 1PE_STUDY_BOLUS_TRACK (Adult)

Source: JHU Protocol
Protocol Range Series Description kV Quality Ref. mAs (Eff.) mAs CAREDose/CAREDose 4D CAREDoseType CTDIvol RotTime Pitch
1PE_STUDY_BOLUS_TRACK (Adult)Topogram Topogram 0.6 T20s 120 0 186 off CAREDose 0 0.500 1.0000000
 PreMonitoring  120 110 60 off CAREDose4D 8.66065 0.330 0.8000000
   PreMonitoring 9.6 B20f        
 I.V. Bolus         
 Monitoring  120 110 60 off CAREDose4D 259.82 0.330 0.8000000
   Monitoring 9.6 B20f        
 PE Study  120 110 225 off CAREDose4D 16.209 0.330 0.80
   PE Study 0.75 B20f -thins        
   PE Study 3.0 B30f        
   PE Study 3.0 B80f - Lung        
Coll. Slice Recon increment Images Kernel Window API Comment1
 0.60 0.60  1 T20s Topogram Body None 
 0.60      None 
  9.60  1 B20f Abdomen  
        
 0.60      None 
  9.60  30 B20f Abdomen  
 0.60      None 
  0.75 0.5 601 B20f PE Window  
  3.00 2.0 151 B30f PE Window  
  3.00 3.0 101 B80f Lung