Siemens Definition 64 Slice MDCT Protocols : Chest : 1PE_STUDY (Adult)

Source: JHU Protocol
Protocol Range Series Description kV Quality Ref. mAs (Eff.) mAs CAREDose/CAREDose 4D CAREDoseType CTDIvol RotTime Pitch

1PE_STUDY (Adult)

Topogram

 Topogram 0.6 T20s

 120

 0

 186

 off

 CAREDose

 0

 0.500

 1.0000000

 

PE Study

 

 120

 110

 200

 off

 CAREDose4D

 14.3948

 0.330

 0.80

 

 

 PE Study 0.75 B20f -thins

 

 

 

 

 

 

 

 

 

 

 PE Study 3.0 B30f

 

 

 

 

 

 

 

 

 

 

 PE Study 3.0 B80f - Lung

 

 

 

 

 

 

 

 

Coll. Slice Recon increment Images Kernel Window API Comment1

 0.60

 0.60

 

 1

 T20s

 Topogram Body

 None

 

 0.60

 

 

 

 

 

 None

 

 

 0.75

 0.5

 601

 B20f

 PE Window

 

 

 

 3.00

 2.0

 151

 B30f

 PE Window

 

 

 

 3.00

 3.0

 101

 B80f

 Lung