Specific Anatomic Region | Thorax | |
Application | Pulmonary embolism | |
Author | Elliot K. Fishman, MD | |
Reference Source | JHU protocol | |
Scanner Used | Siemens Somatom Sensation 16 | |
KV / Effective mAs / Rotation time (sec) | 120 / 130 / 0.5 | |
Detector Collimation (mm) | 0.75 or 1.5 | |
Slice Thickness (mm) | 1 or 2 | |
Feed / Rotation (mm) | 15.0 | |
Kernel | B30f | |
Increment (mm) | 1 or 2 | |
Image Order | cr-ca or ca-cr | |
Oral Contrast | N/A | |
IV Contrast Volume and Type | 100-120 cc | |
Injection Rate | 3-4 cc/ sec | |
Scan Delay (sec) | 23-28 sec | |
3D Technique Used | Real time thin slab VRT or MIP | |
Comment: -2nd Recon: 5x5 lung B80 or 1.25 x 3 provides HRCT to accompany the study. - The images can be acquired in a caudal/cranial direction to limit motion related artifact and artifact related to contrast in SVC and right side of heart. |