| Specific anatomic region | Chest |
| Application | Hi Res Chest (ILD) |
| Author | Elliot K Fishman, MD |
| Reference Source | JHU Protocol |
| Scanner Used | 64 Slice |
| KV/Effective mAs/ Rotation time (sec) | 120/ 125 / 0.5 |
| Detector Collimation (mm) | 0.6 |
| Slice thickness (mm) | 1mm |
| Pitch | 0.9 |
| Kernel | B80f ultra sharp lung window |
| Reconstruction interval | 10mm |
| Image order (acquisition) | Cr-ca |
| Oral contrast | N/A |
| IV contrast volume and type | If needed 120cc Omnipaque 350 |
| Injection rate | 2-2.5cc/sec |
| Scan delay (sec) | 60 sec |
| 3D technique used | VRT & MIP |
| Comment: When 3D is done reconstruction is .75mm at .5 mm intervals. | |