| Specific anatomic region | Cardiac Low HR (Heart Rate) .37 |
| Application | Coronary Arteries - coronary artery disease |
| Author | Elliot K Fishman, MD |
| Reference Source | JHU Protocol |
| Scanner Used | 64 Slice |
| KV/Effective mAs/ Rotation time (sec) | 120/ 850 / .37 |
| Detector Collimation (mm) | .6 |
| Slice thickness (mm) | .75 |
| Pitch | .18 |
| Kernel | B20 |
| Reconstruction interval | .4 |
| Image order (acquisition) | Cr-ca |
| Oral contrast | N/A |
| IV contrast volume and type | 80 cc Visipaque 320 / 40 cc saline |
| Injection rate | 4cc/sec |
| Scan delay (sec) | Test bolus + 6 sec handling time |
| 3D technique used | VRT/MIP |
| Comment: An ideal heart rate for cardiac CT is 60-65 BPM and if too slow can be an issue as well. | |