| Specific anatomic region | Cardiac CTA |
| Application | Test bolus |
| Author | Elliot K Fishman, MD |
| Reference Source | JHU Protocol |
| Scanner Used | 64 Slice |
| KV/Effective mAs/ Rotation time (sec) | 120/ 40 / .5 |
| Detector Collimation (mm) | 10 mm |
| Slice thickness (mm) | 10 mm |
| Pitch | N/A - 0 table feed |
| Kernel | B20 |
| Reconstruction interval | 10mm |
| Image order (acquisition) | N/A |
| Oral contrast | N/A |
| IV contrast volume and type | Visipaque 320 20cc / 40cc Saline flush |
| Injection rate | 4cc/sec |
| Scan delay (sec) | 10 sec and then every 3 sec (10 scans) |
| 3D technique used | N/A |
| Comment: Uses 2 sec cycle time. Test bolus used to determine patient specific delay time for CTA. Delay time + 6 sec is ideal. | |